Lab CMS-2567 ToolkitPlan of correction process and templates Book a free consultation
Home › POC template

CLIA plan of correction template for the CMS-2567

Use one entry per D-tag with seven fields: the tag, action for affected patients, look-back for others, systemic change, monitoring, completion date and responsible job title. The worked example below shows a complete entry for a D5413 temperature monitoring finding.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

CLIA plan of correction template for Form CMS-2567 with example entry
CLIA plan of correction template for Form CMS-2567 with example entry

The template

FieldWhat to write
ID prefix tagThe D-tag exactly as cited (for example D5413)
1. Affected patientsWhat was done for the patients, results or specimens the survey identified: review, corrected reports, notification of ordering providers where results were affected
2. Others potentially affectedHow the lab looked back for other affected patients or records (date range, systems reviewed, method) and what it found and did
3. Systemic changeThe process, procedure, form, LIS setting or staffing change that prevents recurrence, with the procedure name and revision date
4. MonitoringWho checks, how often, what sample size, for how long, the threshold that triggers action, and where results are reported (for example to the laboratory director monthly)
Completion dateA specific date for each correction, realistic and within 12 months of the survey; for condition-level findings, the date already completed
Responsible personA job title, not a personal name

Worked example (illustrative)

Illustrative composite for a D5413 finding (42 CFR 493.1252(b)): reagent refrigerator temperatures not recorded on several days and no range defined.

ID prefix tagD5413
1. Affected patientsThe laboratory director reviewed QC and patient results for all assays using reagents from the unmonitored refrigerator for the dates cited. QC was within limits on each run; no patient results required correction. Review documented and signed.
2. Others potentially affectedThe technical consultant reviewed temperature logs for all refrigerators, freezers, incubators and room temperature for the prior 12 months. Two additional gaps on a freezer log were found; affected materials were evaluated and documented.
3. Systemic changeProcedure "Temperature Monitoring" revised to define acceptable ranges per manufacturer instructions for each unit, required daily recording including weekends, and corrective action steps for out-of-range readings. Continuous electronic monitoring with alarms installed. All testing personnel trained; training records on file.
4. MonitoringSupervisor reviews all temperature records weekly for three months, then monthly. Any missed reading or undocumented corrective action is reported to the director, and results are summarized at the monthly quality meeting for 12 months.
Completion dateA specific date within the 12-month limit
ResponsibleLaboratory supervisor; oversight by laboratory director

Adapting the template

For condition-level tags, change the completion date field to the date the correction was finished and attach or list the evidence that supports your allegation of compliance. For personnel tags, the systemic change is often a qualification file checklist and an appointment letter defining delegated duties under Subpart M.

Need help writing your plan of correction?

Sound Lab Consultants helps laboratory directors work through root causes and write plans of correction and allegations of compliance that answer each D-tag.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

Is there an official CLIA plan of correction template?

No separate template exists. The plan is written in the right-hand column of Form CMS-2567 or an attachment. The structure here follows the four elements in SOM Chapter 6 §6132.2.

Can I attach pages to the CMS-2567?

Labs commonly attach continuation pages when the plan does not fit the column. Keep each entry keyed to its D-tag.

Should evidence be sent with the plan?

For condition-level allegations of compliance, evidence supports credibility. For standard-level plans, keep the evidence ready for review or revisit.