Root cause analysis for CLIA laboratory deficiencies
Root cause analysis asks why the deficient practice happened until you reach a cause in the process: an unclear procedure, a missing check, no time in the schedule, an LIS setting, an unqualified role. Fixing that cause is what makes the systemic-change element of a plan of correction credible.

A practical method
- Write the deficient practice in one sentence, using the surveyor's evidence.
- Map the steps of the process involved, from order to report or receipt to discard.
- Ask "why did this step fail?" five times, writing each answer down.
- Test the cause: would fixing it have prevented every instance in the findings?
- Choose a fix that changes the process, then decide how to prove it works.
Example: the five whys (illustrative)
| Why? | Answer |
|---|---|
| Why were controls past expiration in use? | Nobody removed them when they expired. |
| Why not? | There was no assigned check of open-dated materials. |
| Why not? | The procedure did not define open-vial stability or who checks. |
| Why not? | The procedure was written for the previous control product. |
| Why was it not updated? | Product changes do not trigger procedure review. |
Root cause: product changes do not trigger procedure review. The fix is a change-control step, not just a reminder to check dates.
Causes behind frequently cited tags
- D5413 conditions: undefined ranges, weekend coverage gaps, no escalation for out-of-range readings.
- D5209 competency: no tracking of due dates, assessors unclear on six elements.
- D5401/D5403 procedures: procedures not updated after method or instrument changes.
- D5421 verification: go-live allowed before director approval of verification data.
Need help writing your plan of correction?
Sound Lab Consultants helps laboratory directors work through root causes and write plans of correction and allegations of compliance that answer each D-tag.
Related guides on soundlabconsult.com:
Frequently asked questions
Does CMS require a root cause analysis?
SOM Chapter 6 requires a plan that describes the systemic changes made so the deficient practice does not recur. A root cause analysis is how labs identify which system to change.
Is human error an acceptable root cause?
It is rarely accepted alone. Ask why the process allowed the error and fix that.
Should the root cause analysis be submitted?
Summarize the cause in the plan where it explains the systemic change. Keep the full analysis on file.