Responding to a AST (Aspartate Aminotransferase) Deficiency on the CMS-2567
A strong plan of correction for a AST finding names the fix, shows a look-back for affected patient results, changes the system that failed, assigns monitoring and gives a completion date, with the AST records that prove it attached.
What a AST plan of correction should cover
CMS returns plans that only promise to do better. Plans of correction generally need to explain:
- The fix: what was corrected for AST and when.
- The look-back: how you found other AST results or patients that may have been affected, and what you did about them.
- The system change: the procedure, training or software change that keeps it from happening again.
- The monitor: who checks it, how often, and where it is reported (for example to the laboratory director or QA meeting).
- The date: a realistic completion date for each item.
Evidence to attach for AST
- Daily AST QC records, Levey-Jennings charts, corrective action log, control lot ranges
- Six-month AST calibration verification and reportable range data with director review
- AST PT enrollment, attestation statements, graded reports, investigation of every miss
- AST verification study: accuracy, precision, reportable range, reference interval check, director sign-off
- Critical value list, call log with read-back, report format showing AST reference intervals
- Training and competency records for everyone who runs AST
AST PT and accuracy
AST (Aspartate Aminotransferase) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±15% or ±6 U/L (greater). If the finding involves PT, include the investigation of each unacceptable result and the remedial action taken; routine chemistry rules require that documentation be kept for two years (42 CFR 493.841).
Common root causes behind AST findings
- Hemolysis (in vitro) during collection or transport: red cells contain much more AST than plasma, so even slight hemolysis raises results.
- Delayed separation can raise AST; centrifuge within the time your procedure allows.
- Procedures that do not match what staff actually do.
- No second-level review of QC and calibration verification records.
Related tests
- ALT (Alanine Aminotransferase) (full guide)
- LDH (Lactate Dehydrogenase) (full guide)
- Creatine Kinase (CK) (full guide)
- Bilirubin, Total (full guide)
Full guide: AST (Aspartate Aminotransferase) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Need help writing your plan of correction?
Sound Lab Consultants helps laboratory directors work through root causes and write plans of correction and allegations of compliance that answer each D-tag.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I write a plan of correction for a AST deficiency?
State the correction, the look-back for affected AST results, the system change, the monitoring and a completion date, and attach the supporting records.
What records support a AST plan of correction?
Daily AST QC records, Levey-Jennings charts, corrective action log, control lot ranges; Six-month AST calibration verification and reportable range data with director review; AST PT enrollment, attestation statements, graded reports, investigation of every miss.
Where is the plan of correction written?
On Form CMS-2567, the Statement of Deficiencies and Plan of Correction, or in the format your state agency or accrediting organization specifies.
Sources
- 42 CFR 493.931, Routine chemistry
- Form CMS-2567, Statement of Deficiencies and Plan of Correction
- CMS State Operations Manual, Chapter 6 (Special Procedures for Laboratories)
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493 Subpart H, Participation in Proficiency Testing
- CMS State Operations Manual, Appendix C (Laboratories)