Responding to a Immunoglobulin A (IgA) Deficiency on the CMS-2567
A strong plan of correction for a IgA finding names the fix, shows a look-back for affected patient results, changes the system that failed, assigns monitoring and gives a completion date, with the IgA records that prove it attached.
What a IgA plan of correction should cover
CMS returns plans that only promise to do better. Plans of correction generally need to explain:
- The fix: what was corrected for IgA and when.
- The look-back: how you found other IgA results or patients that may have been affected, and what you did about them.
- The system change: the procedure, training or software change that keeps it from happening again.
- The monitor: who checks it, how often, and where it is reported (for example to the laboratory director or QA meeting).
- The date: a realistic completion date for each item.
Evidence to attach for IgA
- Daily IgA QC records, Levey-Jennings charts, corrective action log, control lot ranges
- Six-month IgA calibration verification and reportable range data with director review
- IgA PT enrollment, attestation statements, graded reports, investigation of every miss
- IgA verification study: accuracy, precision, reportable range, reference interval check, director sign-off
- Critical value list, call log with read-back, report format showing IgA reference intervals
- Training and competency records for everyone who runs IgA
IgA PT and accuracy
Immunoglobulin A (IgA) is a regulated analyte (general immunology, 42 CFR 493.927). CLIA PT acceptance limit: target value ±20%. If the finding involves PT, include the investigation of each unacceptable result and the remedial action taken; routine chemistry rules require that documentation be kept for two years (42 CFR 493.841).
Common root causes behind IgA findings
- Pediatric reference intervals differ sharply by age.
- Nephelometric methods can show antigen-excess effects at very high levels.
- Procedures that do not match what staff actually do.
- No second-level review of QC and calibration verification records.
Related tests
- Immunoglobulin G (IgG) (full guide)
- Immunoglobulin M (IgM)
- Immunoglobulin E (IgE), Total (full guide)
- Total Protein (full guide)
Full guide: Immunoglobulin A (IgA) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.
Need help writing your plan of correction?
Sound Lab Consultants helps laboratory directors work through root causes and write plans of correction and allegations of compliance that answer each D-tag.
Related guides on soundlabconsult.com:
Frequently asked questions
How do I write a plan of correction for a IgA deficiency?
State the correction, the look-back for affected IgA results, the system change, the monitoring and a completion date, and attach the supporting records.
What records support a IgA plan of correction?
Daily IgA QC records, Levey-Jennings charts, corrective action log, control lot ranges; Six-month IgA calibration verification and reportable range data with director review; IgA PT enrollment, attestation statements, graded reports, investigation of every miss.
Where is the plan of correction written?
On Form CMS-2567, the Statement of Deficiencies and Plan of Correction, or in the format your state agency or accrediting organization specifies.
Sources
- 42 CFR 493.927, General immunology
- Form CMS-2567, Statement of Deficiencies and Plan of Correction
- CMS State Operations Manual, Chapter 6 (Special Procedures for Laboratories)
- 42 CFR 493.1256, Control procedures
- 42 CFR 493.1255, Calibration and calibration verification
- 42 CFR 493 Subpart H, Participation in Proficiency Testing
- CMS State Operations Manual, Appendix C (Laboratories)