A strong plan of correction for a BUN finding names the fix, shows a look-back for affected patient results, changes the system that failed, assigns monitoring and gives a completion date, with the BUN records that prove it attached.

Last reviewed Nov 24, 2026 · By the Sound Lab Consultants regulatory team · Sources listed below

What a BUN plan of correction should cover

CMS returns plans that only promise to do better. Plans of correction generally need to explain:

  1. The fix: what was corrected for BUN and when.
  2. The look-back: how you found other BUN results or patients that may have been affected, and what you did about them.
  3. The system change: the procedure, training or software change that keeps it from happening again.
  4. The monitor: who checks it, how often, and where it is reported (for example to the laboratory director or QA meeting).
  5. The date: a realistic completion date for each item.

Evidence to attach for BUN

BUN PT and accuracy

Urea Nitrogen (BUN) is a regulated analyte (routine chemistry, 42 CFR 493.931). CLIA PT acceptance limit: target value ±9% or ±2 mg/dL (greater). If the finding involves PT, include the investigation of each unacceptable result and the remedial action taken; routine chemistry rules require that documentation be kept for two years (42 CFR 493.841).

Common root causes behind BUN findings

Related tests

Full guide: Urea Nitrogen (BUN) testing in the clinical lab: CLIA, PT and QC requirements on soundlabconsult.com, part of the Laboratory Test Compliance Library.

For laboratory compliance and operations only, not medical advice or patient care guidance.

Need help writing your plan of correction?

Sound Lab Consultants helps laboratory directors work through root causes and write plans of correction and allegations of compliance that answer each D-tag.

Book a free consultation

Related guides on soundlabconsult.com:

Frequently asked questions

How do I write a plan of correction for a BUN deficiency?

State the correction, the look-back for affected BUN results, the system change, the monitoring and a completion date, and attach the supporting records.

What records support a BUN plan of correction?

Daily BUN QC records, Levey-Jennings charts, corrective action log, control lot ranges; Six-month BUN calibration verification and reportable range data with director review; BUN PT enrollment, attestation statements, graded reports, investigation of every miss.

Where is the plan of correction written?

On Form CMS-2567, the Statement of Deficiencies and Plan of Correction, or in the format your state agency or accrediting organization specifies.